Health centers in small towns in Bosnia and Herzegovina do not have DAP facility or TLDequipment so that the estimation of entrance surface doses (ESD) could be performed merely from Tube OutputMeasurements. The goal of this work is to evaluate ESD at a local level in X-routine diagnostic radiologyprocedures for standard adult patient, including correlation analysis and comparison the results obtained with therecommended diagnostic reference levels. In a small health center such as the Banovici Health Center in theTuzla Canton, the measurements were focused to high-frequency X-ray examinations: chest (PA) and lumbarspine (AP, Lat). The results obtained indicate that the deviations between ESD values of two radiologistsemployed in the Banovici Health Center are mostly due to very large exposure time during the examinationsdone by the one of radiologists. The value of cross-correlation coefficient between the ESD for chest PAexamination of –0.75 indicates that one data set has a negative linear relationship to another data set. Ahistogram of the result, obtained by the bootstrap procedure establishing the uncertainty of the quantity,represents strong quantitative evidence that both ESD data set are negatively correlated. Also, the correlationcoefficients of 0.12 and 0.41, for the LS examination in PA and LAT projections, respectively, do not indicatestrong linear relation between data set of both radiologists. It was performed ANOVA for comparing the meansof two radiologists data set for the chest AP examination. The very small value of p=1.09*10~(-7) indicates thatdifferences between the two chest ESD results are highly significant. However, p-values of 0.17 and 0.35obtained for LS-examinations in AP and LAT projections, respectively, suggest that the sample means are notsignificantly different. The results indicate that adequate measures should be taken for the benefit of the patients.
机译:波斯尼亚和黑塞哥维那小镇的卫生中心没有DAP设施或TLD 设备,以便仅通过“管输出”即可估算出入口表面剂量(ESD) 测量。这项工作的目的是在X射线常规诊断放射学中评估局部ESD 标准成人患者的程序,包括相关分析和与 推荐的诊断参考水平。在小型保健中心,例如巴努维奇保健中心 在图兹拉州,测量的重点是高频X射线检查:胸部(PA)和腰椎 脊柱(AP,纬度)。获得的结果表明,两位放射科医生的ESD值之间存在偏差 在Banovici保健中心受雇的原因主要是因为检查期间的暴露时间非常长 由一名放射科医生完成。胸部PA的ESD之间的互相关系数值 检查–0.75表示一个数据集与另一数据集具有负线性关系。一种 通过引导程序建立的数量不确定性获得的结果直方图, 表示两个ESD数据集均呈负相关的强有力的定量证据。另外,相关性 分别在PA和LAT投影中进行LS检验的系数0.12和0.41并不表示 两位放射科医师的数据集之间存在很强的线性关系。进行方差分析以比较均值 两名放射科医生的数据集用于胸部AP检查。 p = 1.09 * 10〜(-7)的极小值表示 两次胸部ESD结果之间的差异非常显着。但是,p值为0.17和0.35 分别通过AP和LAT投影中的LS检验获得的结果表明,样本均值不是 明显不同。结果表明应采取适当措施使患者受益。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号