首页> 外文会议>International conference on medical image computing and computer-assisted intervention;MICCAI 2009 >MR to Ultrasound Image Registration for Guiding Prostate Biopsy and Interventions
【24h】

MR to Ultrasound Image Registration for Guiding Prostate Biopsy and Interventions

机译:MR进行超声图像配准以指导前列腺穿刺活检和干预

获取原文

摘要

A method is described for registering preoperative magnetic resonance (MR) to intraoperative transrectal ultrasound (TRUS) images of the prostate gland. A statistical motion model (SMM) of the prostate is first built using training data provided by biomechanical simulations of the motion of a patient-specific finite element model, derived from a preoperative MR image. The SMM is then registered to a 3D TRUS image by maximising the likelihood of the shape of an SMM instance given a voxel-intensity-based feature, which represents an estimate of normal vector at the surface of the prostate gland. Using data acquired from 7 patients, the accuracy of registering T2 MR to 3D TRUS images was evaluated using anatomical landmarks inside the gland. The results show that the proposed registration method has a root-mean-square target registration error of 2.66 mm.
机译:描述了一种用于将术前磁共振(MR)记录到前列腺的术中经直肠超声(TRUS)图像中的方法。首先使用由患者术前MR图像得出的患者特定有限元模型运动的生物力学模拟所提供的训练数据来构建前列腺的统计运动模型(SMM)。然后,在给定基于体素强度的特征的情况下,通过最大化SMM实例形状的可能性来将SMM注册到3D TRUS图像,该特征表示对前列腺表面的法向矢量的估计。使用从7位患者获得的数据,使用腺体内部的解剖标志评估了将T2 MR注册到3D TRUS图像的准确性。结果表明,该配准方法的均方根目标配准误差为2.66 mm。