Image acquisition and real-time processing in region-of-interest fluoroscopy with variation of source-to-image distance

机译:源到图像距离的变化在感兴趣区域荧光检查中进行图像采集和实时处理

获取原文

摘要

Abstract: An automatic ROI tracking system for the application of ROI fluoroscopy is designed. The methods for processing in real- time, generating binary masks, equalization of the display and the scheme for the image analysis to track ROI size are described. The initial results of the system tested on a torso phantom are reported and discussed. The system results in the appearance of an almost completely equalized real-time image and the method can be incorporated into x-ray fluoroscopy units with minimum modification. With the technology, ROI fluoroscopy can be applied to interventional or diagnostic procedures for practical patient dose reduction. !9