A combined cavity ring-down (CRD) and photometry technique is employed to measure the transmittance of optical laser components in a range extending from below 0.01% to over 99.99%. In this combined technique, the conventional photometric configuration is used to measure, by ratioing the transmitted light power to the input power, the transmittance ranging from below 0.01% to over 99% with a typical relative uncertainty below 0.3%, and the CRD configuration is used to measure the transmittance higher than 99% with an uncertainty below 0.01%. Eight test samples with transmittance in the range of nearly 99.99% to approximately 0.013% are experimentally measured. Uncertainties of approximately 0.0001% for the transmittance of 99.9877% and of 0.003% for the transmittance of 0.013% are achieved with respectively the CRD and photometric schemes of a simple experimental apparatus. The experimental results showed that the combined technique is capable of measuring the transmittance of any practically fabricated optical laser components.
机译:采用组合腔响脉冲(CRD)和光度测量技术来测量光激光组分的透射率在低于0.01%至超过99.99%的范围内。在这种组合技术中,通过将传输的光功率与输入功率的比率与输入功率的比率与输入功率的比率相比,使用典型的相对不确定性低于0.3%,并且CRD配置的透射率测量传统的光度配置。用于测量高于99%的透射率,不确定值低于0.01%。实验测量八个透射率的试验近99.99%至约0.013%。对于简单实验装置的CRD和光度法,实现了0.013%的透射率约0.0001%的不确定性,为0.013%的透射率为0.013%。实验结果表明,组合技术能够测量任何实际制造的光激光组件的透射率。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号