Controllability of 1/R control in non-pulsatile total artificial heart

机译:非搏动全人工心脏1 / R控制的可控性

获取原文

摘要

The role of pulsatility in a total artificial heart (TAH) has been a subject of great debate. Additionally, it is unclear whether, in nonpulsatite TAH, a physiological control method such as 1/R control can keep the animal in physiological condition or not. To examine a controllability of 1/R control in non-pulsatile TAH, non-pulsatile flow mode was introduced to the undulation pump TAH (UPTAH) and the hemodynamic parameters in pulsatile and non-pulsatile flow modes with 1/R control was analyzed. The artificial valve was removed from UPTAH and the goat's natural heart was replaced with it. The UPTAH was first driven with a pulsatile flow mode and 1/R control was applied within one week after surgery. The pulsatile flow mode was switched to a non-pulsatile flow mode on the day 2 weeks after surgery. General condition did not change before and after the flow mode twitching. Cardiac output and arterial pressure changed with the physiological condition of the goat in also non-pulsatile flow mode, which seemed not different from that in pulsatile flow mode. However, the sucking effect of atria was very significant in nonpulsatile mode. Although 1/R control could keep the animal in physiological condition in a non-pulsatile flow mode, nonpulsatile TAH was considered to be inadequate for 1/R control to stabilize atrial pressures in the present circumstances or some pulsatility would be necessary to avoid fatal sucking effect and to ensure sufficient inflow condition.
机译:搏动性在全人工心脏(TAH)中的作用一直是一个充满争议的话题。另外,还不清楚在非脉冲石TAH中,诸如1 / R控制的生理控制方法能否使动物保持生理状态。为了检查非搏动性TAH中1 / R控制的可控性,将非搏动性流动模式引入了起搏泵TAH(UPTAH),并分析了具有1 / R控制的搏动性和非搏动性流动模式中的血液动力学参数。从UPTAH上取下了人工瓣膜,并用山羊的天然心脏代替了它。首先以脉动血流模式驱动UPTAH,并在手术后一周内应用1 / R控制。在手术后两周的那天,脉动血流模式切换为非脉动血流模式。流动模式抽动前后,一般情况没有改变。在非脉动血流模式下,心输出量和动脉压也随山羊的生理状况而变化,这似乎与脉动血流模式下没有差异。但是,在非搏动模式下,心房的吮吸效果非常显着。尽管1 / R控制可以使动物保持在非脉动血流模式下的生理状态,但是在当前情况下,非脉冲TAH被认为不足以进行1 / R控制来稳定心房压,或者为了避免致命性吮吸需要一定的脉动性效果并确保充足的流入条件。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号