Database MAMOLIT for mammography screening patients in Lithuania

机译:数据库MAMOLIT用于在立陶宛进行乳房X线筛查的患者

获取原文

摘要

Patient doses in mammography screening examinations in Lithuania have been investigated within a period between 2003 and 2009. In vivo dose measurements were performed using thermoluminescence dosimetry (TLD) technique. Interactive database MAMOLIT was created, which allowed calculating of patient's doses, documenting of measured and calculated doses and other parameters of the mammography screening examination, following up of the frequency of patient related X-ray examinations, and analysing of the collected data from different points of view. Reference level of 2.4 mGy of the average glandular dose was set for mammography screening in Lithuania on the base of the obtained results. Present paper summarises gathered experience and outlines some problems which may occur when performing dose measurements during mammography screening examinations of patients.
机译:在2003年至2009年之间的一段时间内,对立陶宛的乳房X线照片筛查检查中的患者剂量进行了研究。体内剂量的测量使用热发光剂量法(TLD)进行。创建了交互式数据库MAMOLIT,该数据库允许计算患者的剂量,记录测量和计算出的剂量以及乳房X线照片筛查检查的其他参数,跟进与患者相关的X射线检查的频率以及分析从不同点收集的数据观点。根据获得的结果,为立陶宛的乳房X线检查筛查设定了2.4 mGy的平均腺体剂量参考水平。本文总结了总结的经验,并概述了在患者的乳房X光检查中进行剂量测量时可能出现的一些问题。