首页> 外文会议>World congress on medical physics and biomedical engineering;International congress of the IUPESM >Feasibility of Polycrystalline Alanine-in-Glass Tubes as Gamma-Ray Dosimeters
【24h】

Feasibility of Polycrystalline Alanine-in-Glass Tubes as Gamma-Ray Dosimeters

机译:多晶丙氨酸玻璃管作为伽马射线剂量计的可行性

获取原文

摘要

This study examines the feasibility of utilizing pure polycrystalline alanine-in-glass tubes as commercial dosimeters. The design of the dosimeters avoids preparation complications such as mortaring, sieving, and/or binder addition. It also reduces several fabrication techniques of alanine dosimeters and hence it allows other laboratories to conduct similar electron paramagnetic resonance (EPR) measurements. The dosimeters were irradiated with gamma rays ranging from 0 to 20 Gy. The dosimeters were found to be more sensitive to low radiation doses than other types of alanine-with-binder dosimeters. They exhibited a linear response in the dose range from 0.1 to 20 Gy. These positive properties favor the pure polycrystalline alanine-in-glass tube as a radiation dosimeter for low gamma irradiation doses.
机译:这项研究检验了使用纯多晶丙氨酸玻璃管作为商业剂量计的可行性。剂量计的设计避免了准备工作的复杂性,例如打浆,筛分和/或添加粘合剂。它还减少了丙氨酸剂量计的几种制造技术,因此,它允许其他实验室进行类似的电子顺磁共振(EPR)测量。用0至20 Gy的伽马射线辐照剂量计。发现该剂量计比其他类型的带粘合剂的丙氨酸剂量计对低辐射剂量更敏感。它们在0.1至20 Gy的剂量范围内表现出线性响应。这些积极的特性有利于纯多晶玻璃丙氨酸管作为低剂量伽马射线辐射剂量的辐射剂量计。