首页> 外文会议>World congress on medical physics and biomedical engineering;International congress of the IUPESM >Pulsatile Flow Phantoms for Quality Control of Doppler Ultrasound Imaging
【24h】

Pulsatile Flow Phantoms for Quality Control of Doppler Ultrasound Imaging

机译:多普勒超声成像质量控制的脉冲流幻影

获取原文
获取外文期刊封面目录资料

摘要

In this work it is proposed the development of phantoms to allow de quality control of ultrasound equipments functioning on the Doppler mode with national technology and accessible cost. The device consists of a hydraulic circuit with a peristaltic pump, mixture simulating human tissue, tubes simulating arteries and blood-mimicking fluid. The peristaltic digital pump made the simulation of cardiac beats possible, operating at flow rates of 45 to 150 ml/min and pulse rates of 60 to 200 beats/min. The blood-mimicking fluid simulated the human red cells as particles of PVC powder with approximately 64 |μm diameter, distilled water, glycerin, detergent and dextran.
机译:在这项工作中,提出了幻象的开发,以允许以国家技术和可负担的成本对在多普勒模式下工作的超声设备进行质量控制。该设备由带有蠕动泵的液压回路,模拟人体组织的混合物,模拟动脉的管和模拟血液的流体组成。蠕动数字泵使心跳的模拟成为可能,在45至150 ml / min的流速和60至200次/ min的脉搏速率下运行。模仿血液的流体将人类红细胞模拟为直径约为64μm的PVC粉,蒸馏水,甘油,洗涤剂和右旋糖酐的颗粒。