首页> 外文会议>European Signal Processing Conference >PRELIMINARY RESULTS ON NOISE REDUCTION USING STACKGRAMS FOR (LOW-DOSE) X-RAY CT SINOGRAMS
【24h】

PRELIMINARY RESULTS ON NOISE REDUCTION USING STACKGRAMS FOR (LOW-DOSE) X-RAY CT SINOGRAMS

机译:使用StackGrams(低剂量)X射线CT SINOGRAMS的降噪结果

获取原文

摘要

In low-dose X-ray computed tomography (CT), it is necessary to reduce the excessive X-ray quantum noise of the acquired data. This is the case especially when the filtered back-projection method is used for image reconstruction from sinogram data. In this study, we compare the conventional radial filtering direction of the sinogram to the recently introduced angular stackgram filtering technique. Our experimental tests were performed on simulated low-dose CT data. We modelled the X-ray quantum noise using a Gaussian distribution with a signal dependent variance for numerical phantoms simulating stop-and-shoot CT acquisitions. The noise model has been reported elsewhere. This experimentally verified model describes calibrated projection data, i.e. the acquired data after CT system calibration (including logarithmic transform, beam hardening, etc.). For our quantitative tests, we employed both linear Gaussian low-pass filters and non-linear L-filters with Gaussian weights. The stackgram technique provides the best resolution-noise trade-off in comparison with the radial sinogram filtering method, and thus a potential filtering approach to low-dose CT sinograms.
机译:在低剂量X射线计算机断层扫描(CT)中,需要降低所获取的数据的过度X射线量子噪声。这是这种情况,特别是当过滤的后投影方法用于来自Sinogram数据的图像重建时。在这项研究中,我们将常规径向过滤方向与最近引入的角度堆叠滤波技术进行比较至最近引入的角度堆射技术。我们的实验测试是对模拟的低剂量CT数据进行的。我们使用高斯分布模拟了X射线量子噪声,其具有信号依赖性方差,用于模拟停止和拍摄CT采集的数值幽灵。噪声模型已在其他地方报告。该实验验证的模型描述了校准投影数据,即CT系统校准(包括对数变换,光束硬化等)之后所获得的数据。对于我们的定量测试,我们使用具有高斯重量的线性高斯低通滤波器和非线性L滤波器。与径向铭文滤波方法相比,StackGram技术提供了最佳分辨率噪声折议,从而提供了低剂量CT SINOGROM的潜在过滤方法。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号