We present an 180nm HV CMOS IC for concurrent neural stimulation and recording that combines 64 low-noise recording front-ends and 4 independent stimulators multiplexed to any of the 64 channels. The stimulators have 5mA peak current, 12V compliance and dynamic power management to maximize efficiency. Co-design of the stimulation and recording subsystems resulted in 100mV of recording linear range, 70nV/rtHz noise, and a rapid 1ms (single-sample) artifact recovery during stimulation.
机译:我们介绍了一个180nm的HV CMOS IC,用于并发神经刺激和记录,该记录结合了64个低噪声记录前端和4个独立刺激器多路复用到64个通道中的任何一个。刺激器具有5mA峰值电流,12V合规性和动态电源管理,以最大限度地提高效率。刺激和记录子系统的共同设计导致记录线性范围,70nV / Rthz噪声和刺激期间的快速1ms(单样本)伪影恢复的100mV。