3D model of CZT /spl gamma/-ray spectrometer detector: validation on a monolithic pixelated detector and on a high energy spectrometric probe

机译:CZT / spl伽玛射线光谱仪探测器的3D模型:在单块像素化探测器和高能谱仪上进行验证

获取原文

摘要

A fully three-dimensional model of a semiconductor gamma-ray detector is presented. The model takes into account the gamma-ray and charge collection physical phenomena involved in the detection process and models the readout electronic response and noise. The model successively involves the Monte Carlo simulation of the photon transport, the finite element transient computation of the adjoint transport equation, and the electronic signal processing including an accurate noise model. The simulation is possible because the solution of the adjoint equation allows us to obtain a continuous mapping of the induced charge collection efficiency (at any time and any interaction point) with a single 3D transient finite element computation. The simulation outputs are pulse height spectra and biparametric (rise time versus pulse height) spectra. The model has been validated on two applications. The first application is a CdZnTe monolithic pixelated detector at medical energies. The detector dimension is 10x10x5 mm, with a pixel size of 2x2 mm, and a 2.5 mm pitch. The second application is a gamma-ray spectrometric single-channel probe for 200 keV to 2 MeV photons combining a small anode and the capacitive Frisch grid structure. The detector dimension is 10x10x9 mm. Simulated and experimental results are compared. In both cases, the comparison between simulated and experimental spectra is qualitatively good for pulse height spectra as well as biparametric spectra.
机译:提出了半导体伽马射线探测器的完整三维模型。该模型考虑了检测过程中涉及的伽马射线和电荷收集物理现象,并对读取的电子响应和噪声进行了建模。该模型依次涉及光子输运的蒙特卡洛模拟,伴随输运方程的有限元瞬态计算以及包括精确噪声模型的电子信号处理。该模拟是可行的,因为伴随方程的解使我们能够通过单个3D瞬态有限元计算获得感应电荷收集效率(在任何时间和任何相互作用点)的连续映射。模拟输出是脉冲高度谱和双参数(上升时间与脉冲高度)谱。该模型已在两个应用程序上得到验证。第一个应用是医疗能量下的CdZnTe整体式像素化探测器。检测器尺寸为10x10x5 mm,像素尺寸为2x2 mm,间距为2.5 mm。第二个应用是用于200 keV至2 MeV光子的伽马射线光谱单通道探测器,该探测器结合了小阳极和电容性Frisch网格结构。检测器尺寸为10x10x9 mm。比较了模拟结果和实验结果。在这两种情况下,对于脉冲高度谱以及双参数谱,模拟谱图和实验谱图之间的比较在质量上都很好。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号