Comparison of Current Densities measured in a Pig Heart during States of Ventricular Fibrillation and Post-mortem for different Defibrillation Electrode Positions

机译:不同除颤电极位置在心室颤动和验尸期间在猪心脏中测量的电流密度的比较

获取原文

摘要

Current density imaging (CDI) is an MRI technique used to quantitatively measure current density vectors in biological tissue. A fast CDI sequence was developed that can image the whole body of a 4 kg pig in about 15 minutes. A state of ventricular fibrillation (VF) can be sustained for nearly 30 minutes allowing two complete CDI scans of the same subject. A single parameter, i.e. electrode position, is adjusted between the two scans for comparative analysis. This study compares the current density vector directions and current density magnitudes measured for two typical electrode positions, i.e. apex anterior (AA) and apex posterior (AP). The comparative experiment is repeated on the same subjects for states of immediate post-mortem and one hour post-mortem. Further, the acquired vector datasets are used to compute conductivity images of the heart
机译:电流密度成像(CDI)是一种MRI技术,用于定量测量生物组织中的电流密度矢量。快速的CDI序列得以开发,可以在大约15分钟内对4千克猪的整个身体进行成像。心室颤动(VF)的状态可以维持近30分钟,从而允许对同一受试者进行两次完整的CDI扫描。在两次扫描之间调整单个参数,即电极位置,以进行比较分析。这项研究比较了两个典型电极位置(即先尖(AA)和后尖(AP))的电流密度矢量方向和电流密度大小。对同一受试者重复进行即时死后和死后一小时状态的比较实验。此外,所获取的矢量数据集用于计算心脏的电导率图像