首页> 外文会议> >A catheter guiding method for ablative therapy of cardiac arrhythmias by application of an inverse solution to body surface electrocardiographic signals
【24h】

A catheter guiding method for ablative therapy of cardiac arrhythmias by application of an inverse solution to body surface electrocardiographic signals

机译:通过对体表心电图信号应用逆溶液消融治疗心律失常的导管引导方法

获取原文

摘要

This study investigates the effectiveness of an algorithm in guiding the ablative therapy of cardiac arrhythmias. In this process both the site of origin of the arrhythmia and current pulses delivered from the tip of the catheter are modeled with a single equivalent moving dipole (SEMD) model in an infinite homogeneous volume conductor. Because of the lack of accurate torso geometry, the algorithm introduces systematic error in the estimated compared to the true dipole position. Computer simulations have shown that this systematic error has minor influence in guiding the tip of the catheter to the site of the origin of the arrhythmia. In this study, to further investigate the accuracy of this method, we conducted experiments using a 0.9% saline volume-conductor. Our results demonstrate that the final distance between the catheter tip and the target site is less than 3 mm and thus the lack of accurate torso geometry may have minor influence in the catheter guidance.
机译:本研究研究了算法在引导心律失常的烧蚀治疗方面的有效性。在该过程中,从导管的尖端输送的心律失常的原点和电流脉冲都用单一等效移动的偶极(SEMD)模型在无限均匀的体积导体中进行建模。由于缺乏准确的躯干几何形状,算法在与真正的偶极位置相比估计中引入了系统误差。计算机模拟表明,这种系统误差对导管的尖端引导到心律失常的起源部位的影响。在该研究中,为了进一步研究该方法的准确性,我们使用0.9%盐水体积导体进行实验。我们的结果表明,导管尖端和靶位点之间的最终距离小于3mm,因此缺乏精确的躯干几何形状可能对导管引导有微小的影响。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号