Measured accuracy and precision in quantitative SPECT imaging with iodine-123

机译:用碘123测定定量SPECT成像中的准确性和精密度

获取原文
获取外文期刊封面目录资料

摘要

This study evaluated /sup 123/I SPECT volume and activity quantitation in terms of accuracy and precision using experimental phantom data. Two reconstruction methods were considered. The filtered backprojection (FB) method with 2D Metz pre-filtering and single iteration Chang attenuation compensation was compared with an iterative ML-EM method that used Gaussian post-filtering and incorporated the 3D divergent detector response and attenuation in the detection kernel. Both methods were examined over a range of resolution and noise characteristics. An ensemble of 21 SPECT data sets were acquired of a cylindrical phantom that contained spheres of size 21.4, 12.0, and 5.7 ml, filled with high purity /sup 123/I in a uniform background (5:1 uptake ratio). Each data set was reconstructed by both methods, and sphere volume and activity measurements were obtained from the reconstructions using a semi-automatic method that defines the region-of-interest based on the intensity gradient in the area of increased /sup 123/I concentration. The mean and standard deviation of the 21 measurements provided accuracy and precision indices, respectively. For the 21.4 ml sphere, the measured volume with either reconstruction method was accurate to within 5% of the true volume with a standard deviation of approximately 8% of the true volume. For the 12.0 ml sphere the measured volume with the iterative method exhibited substantially smaller standard deviation for equal accuracy compared with the FB method. Sphere activity was underestimated with both methods due to spatial resolution effects. The 5.7 ml sphere was generally below the size limit for volume and activity quantitation by these methods.
机译:这项研究使用实验体模数据在准确性和精密度方面评估了/ up 123 / I SPECT的体积和活性定量。考虑了两种重建方法。将具有2D Metz预滤波和单迭代Chang衰减补偿的滤波反投影(FB)方法与使用高斯后滤波并将3D发散检测器响应和衰减合并到检测内核中的迭代ML-EM方法进行了比较。在分辨率和噪声特性的范围内检查了这两种方法。获得了21个SPECT数据集的集合,其中包含一个圆柱体模,该体模包含大小为21.4、12.0和5.7 ml的球体,并在均匀的背景下(吸收比为5:1)填充了高纯度/ sup 123 / I。通过这两种方法重建每个数据集,并使用半自动方法从重建中获得球体体积和活动度测量值,该方法基于增加的/增加的123 / I浓度区域中的强度梯度来定义感兴趣区域。这21次测量的平均值和标准差分别提供了准确性和精度指标。对于21.4 ml的球体,用任何一种重建方法测得的体积均精确到真实体积的5%以内,标准偏差约为真实体积的8%。对于12.0 ml的球体,与FB方法相比,用迭代法测得的体积显示出较小的标准偏差,以实现相同的精度。由于空间分辨率的影响,两种方法都低估了球体的活动性。 5.7 ml的球体通常低于通过这些方法进行体积和活性定量的大小限制。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号