Mechanical effect of contrast injection on coronary artery diameter during angiography

机译:造影剂造影剂注射对冠状动脉直径的机械作用

获取原文

摘要

The authors performed an angiographic study in six healthy anaesthetized pigs to investigate possible mechanical effects of contrast injection on coronary artery pressure and diameter during the period of image acquisition. Angiograms were acquired at 30 f/sec for repeated injections of 8 ml of contrast material by hand and with a power injector at rates of 2, 4, and 6 ml/sec. A 1.5 F micro-tip pressure transducer was used to measure intracoronary pressure at 1, 2, and 4 cm downstream from the injection site for each injection rate. The results show a progressive increase in intracoronary pressure with increasing rate of injection (p>0.005), ranging from 3 to 117 mmHg. Corresponding diameters were distended by 0.02 to 0.6 mm (p>0.05) during the injection, as evaluated by quantitative coronary angiography (QCA). These mechanical effects were related to the mean coronary pressure prior to injection (p>0.02) in a manner consistent with the nonlinear compliance of the arterial wall. The authors conclude that coronary dimensions obtained from QCA critically depend on the mode of contrast injection, which may represent a previously unrecognized source of error in the analysis and interpretation of clinical angiographic studies.
机译:作者对六只健康麻醉的猪进行了血管造影研究,以研究在图像采集期间造影剂注射对冠状动脉压力和直径的可能的机械作用。以30 f / sec的速度采集血管造影照片,以手动方式和电动注射器以2 ml,4 ml和6 ml / sec的速率重复注射8 ml造影剂。对于每个注射速率,使用1.5 F微尖端压力传感器测量注射部位下游1、2和4 cm处的冠状动脉内压力。结果显示,随着注射速率的增加,冠状动脉内压逐渐升高(p> 0.005),范围从3至117 mmHg。通过定量冠状动脉造影(QCA)评估,注射过程中相应的直径扩大了0.02至0.6 mm(p> 0.05)。这些机械作用与注射前的平均冠状动脉压力有关(p> 0.02),其方式与动脉壁的非线性顺应性一致。作者得出的结论是,从QCA获得的冠状动脉尺寸严重取决于造影剂注射的方式,这可能代表了先前在临床血管造影研究的分析和解释中无法识别的错误来源。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号