首页> 外文会议> >Design and characterization of a 10 cm annular array transducer for high intensity focused ultrasound (HIFU) applications
【24h】

Design and characterization of a 10 cm annular array transducer for high intensity focused ultrasound (HIFU) applications

机译:用于高强度聚焦超声(HIFU)应用的10 cm环形阵列换能器的设计和表征

获取原文

摘要

A large transducer to generate high intensity focused ultrasound (HIFU) for surgery via tissue ablation is described. Designed to address the need for therapeutic applications into deep tissue, the device operates at 1.25 MHz. In order to increase focus-to-entrance intensity ratios, the design was optimized to have a natural focal distance of 10 cm at f/1.0. Variable focus is achieved by phasing the array, and particular care was exercised to maximize the range of focal distances that can be achieved with a limited number of amplifiers and elements, using a constant-phase delay design. Design approach and tradeoffs are discussed, with emphasis on comparing theoretical calculations with test performance.
机译:描述了一种大型换能器,该换能器通过组织消融产生用于手术的高强度聚焦超声(HIFU)。该设备旨在满足深层组织治疗应用的需求,其工作频率为1.25 MHz。为了增加焦点与入口的强度比,对设计进行了优化,使其在f / 1.0时具有10 cm的自然焦距。通过定相阵列可实现可变聚焦,并且使用恒定相位延迟设计,要格外小心,以使有限数量的放大器和元件可以实现的焦距范围最大化。讨论了设计方法和折衷方案,重点是将理论计算与测试性能进行比较。