Studies on soft tissue pressure distribution in the arm during non-invasive blood pressure measurement

机译:无创血压测量过程中手臂软组织压力分布的研究

获取原文
获取外文期刊封面目录资料

摘要

The transmission of the pressure applied by means of an external cuff to the soft tissues of the arm was analyzed in order to understand significant causes of errors. The problem was studied with a finite element model that showed that the periarterial pressure depends upon several factors, such as the cuff size, the soft tissue elastic properties of the arm and the artery position. Experimental trials were performed with physical models to verify the finite element model and characterize the behavior of the cuff. The results indicate that the proposed model may contribute a simple new tool that is able to describe the tissue and cuff behavior with sufficient accuracy to be useful in non-invasive arterial pressure measurements.
机译:分析了通过外部袖带施加到手臂软组织的压力,以了解导致错误的重要原因。用有限元模型研究了该问题,该模型表明动脉周围压力取决于几个因素,例如袖带尺寸,手臂的软组织弹性和动脉位置。用物理模型进行了试验,以验证有限元模型并表征袖带的行为。结果表明,所提出的模型可能有助于简单的新工具,该工具能够以足够的精度描述组织和袖带的行为,以用于非侵入性动脉压测量。