Removal of harmonic artifacts from synchrotron radiation coronary angiograms

机译:从同步辐射冠状血管造影图中去除谐波伪影

获取原文
获取外文期刊封面目录资料

摘要

Intravenous coronary angiograms of human subjects are now available from the National Synchrotron Light Source (NSLS). The high (4.4-T) magnetic field used under normal operation of the superconducting wiggler magnet on the X-17 beam line at NSLS yields a high critical energy (18.7 keV) and a large harmonic contamination of the desired X-ray flux at the K-edge of iodine (33.169 keV). The 66- and 99-keV contamination is a factor of 2.29 and 5.32 worse than for the human angiograms taken at the Stanford Synchrotron Radiation Laboratory, where a critical energy of 13.0 keV was obtained at beam line IV-2. This increased harmonic contamination produces severe artifacts in the images, especially where the patient's ribs cross the left ventrical and aorta, which are filled with contrast agent. An algorithm was developed to remove these artifacts by solving the two simultaneous transcendental equations for the fractional X-ray transmissions through iodine and through water. The parameters are derived from measurements on a Lucite step-wedge phantom. Human angiograms taken at 4.4 T are shown before and after the correction algorithm is applied. These corrected angiograms are compared to uncorrected angiograms taken at 3.0 T.
机译:现在可以从国家同步加速器光源(NSLS)获得人类受试者的静脉冠状动脉造影照片。在NSLS的X-17射线线上超导摆动磁体正常工作下使用的高(4.4-T)磁场会产生较高的临界能量(18.7 keV),并在所需的X射线通量下产生较大的谐波污染。碘的K边缘(33.169 keV)。 66 keV和99 keV的污染比斯坦福同步加速器辐射实验室拍摄的人体血管造影照片差2.29和5.32倍,后者在束线IV-2处获得了13.0 keV的临界能量。这种增加的谐波污染会在图像中产生严重的伪影,尤其是在患者的肋骨穿过左心室和主动脉的地方,这些肋骨充满了造影剂。开发了一种算法,通过求解两个同时通过的先验方程来去除碘和水的X射线透射分数,从而消除这些伪影。这些参数是从Lucite台阶楔体模上的测量得出的。显示了在应用校正算法之前和之后在4.4 T下拍摄的人体血管造影照片。将这些校正后的血管造影照片与在3.0 T下拍摄的未校正后的血管造影照片进行比较。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号