Surface-mounted dry electrode and analog-front-end systems for physiological signal measurements

机译:表面安装的干电极和模拟前端系统,用于生理信号测量

获取原文

摘要

In this paper, we present a new analog-front-end system including surface-mounted dry electrode by MEMS dry electrode (MDE) and instrumentation amplifier by chopper-stabilized differential difference amplifier (CHDDA) for physiological signal recording applications. Comparing to traditional electrodes, proposed MDE shows its superior advantages for low electrode skin interface impedance and need not to use electrolytic gel. Moreover, the measured results of the presented CHDDA demonstrated the low power, denoise capability for physiological signal recording. The full AFE system is 4.5*7cm2 in size, weight 35g and can be operated for 200Hr by 2 AAA batteries. Finally, the actual 4 channel AFE recording of EEG, EOG, ECG and EMG proves the practicability of the proposed system.
机译:在本文中,我们提出了一种新的模拟前端系统,该系统包括用于MEMS干电极(MDE)的表面安装干电极和用于斩波器的差分差动放大器(CHDDA)的仪表放大器,用于生理信号记录应用。与传统电极相比,建议的MDE显示出其低电极皮肤界面阻抗的优越优势,并且无需使用电解凝胶。而且,所提出的CHDDA的测量结果证明了用于生理信号记录的低功率,降噪能力。完整的AFE系统尺寸为4.5 * 7cm 2 ,重量35g,可通过2节AAA电池运行200小时。最后,实际的4通道AFE记录的EEG,EOG,ECG和EMG证明了该系统的实用性。