A dose assessment study was carried out to determine the likely annual equivalent doses received by variousrnparts of a cardiologist’s body. High sensitivity GR-200 thermoluminescent dosemeters were attached torncardiologists’ foreheads, little fingers, wrists, elbows, knees and ankles. Three common cardiology proceduresrnwere investigated, namely, percutaneous transluminal coronary angioplasty (PTCA), permanent pacemakerrninsertion (PPM) and left heart catheterisation (LHC). Dose monitoring was done on a case-by-case basis. Data onrnten cases of each procedure were gathered. The projected annual equivalent doses were computed by averagingrnthe ten doses measured at each site for each examination type and finding out from the cardiologists how manyrncases of PTCA, PPM and LHC they do in a year. Results in this study show that for the lens of the eye, thernprojected annual equivalent dose is below 10 mSv and for the other body parts, it is below 100 mSv per year. Thernstudy demonstrated that the methodology used can help to optimise radiation protection in diagnostic radiology.
机译:进行了剂量评估研究,以确定心脏病专家身体各个部位可能接受的年度等效剂量。高灵敏度GR-200热发光剂量计安装在心脏病专家的前额,小指,手腕,肘部,膝盖和脚踝上。研究了三种常见的心脏病程序,即经皮腔内冠状动脉成形术(PTCA),永久性起搏器插入(PPM)和左心导管插入(LHC)。剂量监测是逐案进行的。收集每个程序的病例数据。通过平均每种检查类型在每个站点测量的十个剂量,并从心脏病专家那里找出他们一年中发生的PTCA,PPM和LHC病例数的平均值,可以计算出预计的年度等效剂量。这项研究的结果表明,对于晶状体,预计的年等效剂量低于10 mSv,而对于其他身体部位,则每年低于100 mSv。研究表明,所使用的方法可以帮助优化放射诊断中的辐射防护。