首页> 外文期刊>Radiation Protection Dosimetry >CORRELATION OF EYE LENS DOSES AND PERSONAL DOSE EQUIVALENT MEASURED ON THE ARM OF INTERVENTIONAL CARDIOLOGISTS FOR A RETROSPECTIVE ASSESSMENT OF DOSES TO OPERATORS' EYE LENS
【24h】

CORRELATION OF EYE LENS DOSES AND PERSONAL DOSE EQUIVALENT MEASURED ON THE ARM OF INTERVENTIONAL CARDIOLOGISTS FOR A RETROSPECTIVE ASSESSMENT OF DOSES TO OPERATORS' EYE LENS

机译:在介入心脏病学家手臂上测量的眼镜剂量和个人剂量等同物的相关性对操作员眼镜透镜的回顾性评估

获取原文
获取原文并翻译 | 示例
           

摘要

Coefficients converting the readings of the whole body dosemeter worn on the left arm to eye lens doses were determined by analysing the correlations between H_p(10) and H_p(3) values. Doses were measured on a phantom for specific C-arm projections typically used during CA/PCI procedures. In order to estimate the cumulative eye lens doses, conversion coefficients were then applied to the dose records of interventional cardiologists collected in the database of dosimetry service between the years 1995 and 2009. The H_p(10) to H_p(3) conversion coefficients are 0.29 (CV = 34%) and 0.17 (CV = 42%) for left and right eye lens, respectively. However, they can vary from one laboratory to another depending on working technique. From among 61 interventional cardiologists, none exceeded the threshold dose of 0.5 Gy for eye lens opacities. However, 44% of interventional cardiologists were likely to exceed the annual limit of 20 mSv for the most exposed eye at least once in the analysed time period.
机译:通过分析H_P(10)和H_P(3)值之间的相关性来决定转换在左臂上佩戴在左臂上的整体剂量表的读数。在通常在CA / PCI程序期间使用的特定C臂突起的幻像测量剂量。为了估计累积的眼睛透镜剂量,然后将转化系数应用于1995年和2009年之间的剂量测定服务数据库中收集的介入心脏病学家的剂量记录。H_P(10)到H_P(3)转换系数为0.29 (CV = 34%)和0.17(CV = 42%)分别用于左眼镜头。然而,根据工作技术,它们可以从一个实验室变化。从61名介入的心脏病学家中,没有超过眼睛透镜不透明度0.5GY的阈值剂量。然而,44%的介入心脏病学家可能超过在分析的时间段内至少一次暴露于最暴露的眼睛的年限20 msv。

著录项

  • 来源
    《Radiation Protection Dosimetry》 |2020年第3期|271-278|共8页
  • 作者单位

    Radiation Protection Department Nofer Institute of Occupational Medicine 8 St. Teresa Str Lodz 91-348 Poland;

    Radiation Protection Department Nofer Institute of Occupational Medicine 8 St. Teresa Str Lodz 91-348 Poland;

    Radiation Protection Department Nofer Institute of Occupational Medicine 8 St. Teresa Str Lodz 91-348 Poland;

  • 收录信息 美国《科学引文索引》(SCI);美国《化学文摘》(CA);