【24h】

Development of a MRI-Safe Piezo Actuator Drive (PAD)

机译:开发MRI安全压电执行器驱动器(PAD)

获取原文
获取原文并翻译 | 示例

摘要

The Piezo Actuator Drive (PAD) presented by Kappel and Rucha on the Actuator 2006 exhibits a combination of characteristics that makes it ideally suitable for robotics and positioning tasks. The experiments showed that the torque of the PAD and its general function are not considerably affected by the 1.5 Tesla magnetic field of a Magnetic Resonance Imaging device (MRI). Replacing the ferromagnetic components by non-magnetic high-grade steel with the same elastic modulus does neither significantly affect the torque nor the other mechanical characteristics. The attraction of the magnetic field on the PAD is 50 percent of its gravity.
机译:Kappel和Rucha在执行器2006上展出的压电执行器驱动器(PAD)具有多种特性,使其非常适合机器人技术和定位任务。实验表明,磁共振成像设备(MRI)的1.5 Tesla磁场对PAD的扭矩及其一般功能没有太大影响。用具有相同弹性模量的非磁性高级钢代替铁磁组件不会显着影响扭矩或其他机械特性。磁场在PAD上的吸引力为其重力的50%。