首页> 外文会议>International Conference on Medical Biometrics(ICMB 2008); 20080104-05; Hong Kong(CN) >Research on Gesture Definition and Electrode Placement in Pattern Recognition of Hand Gesture Action SEMG
【24h】

Research on Gesture Definition and Electrode Placement in Pattern Recognition of Hand Gesture Action SEMG

机译:手势动作SEMG模式识别中的手势定义和电极位置研究

获取原文
获取原文并翻译 | 示例

摘要

The goal of this study is to explore the effects of electrode placement on the hand gesture pattern recognition performance. We have conducted experiments with surface EMG sensors using two detecting electrode channels. In total 25 different hand gestures and 10 different electrode positions for measuring muscle activities have been evaluated. Based on the experimental results, dependencies between surface EMG signal detection positions and hand gesture recognition performance have been analyzed and summarized as suggestions how to define hand gestures and select suitable electrode positions for a myoelectric control system. This work provides useful insight for the development of a medical rehabilitation system based on EMG technique.
机译:这项研究的目的是探索电极放置对手势模式识别性能的影响。我们使用两个检测电极通道对表面肌电传感器进行了实验。总共评估了25种不同的手势和10种不同的电极位置以测量肌肉活动。基于实验结果,分析了表面肌电信号检测位置与手势识别性能之间的相关性,并总结了如何定义手势和选择适合肌电控制系统的电极位置的建议。这项工作为基于EMG技术的医疗康复系统的开发提供了有用的见识。