【24h】

Newly Established Hydrometer Calibration Set Up at UM

机译:UM新建的比重计校准设置

获取原文
获取原文并翻译 | 示例

摘要

Hydrometer calibration could be performed by 'Cuckow Method' in which just a single liquid is sufficient for the calibration of the whole scale. Hydrometers are weighed first in air and then in a reference liquid in this method. The density of the reference liquid is well known and it is possible to determine the associated density value of the graduation mark by making use of weighing values and the density value of liquid in related formulas. The entire scale of Hydrometer can be calibrated in this way by immersing the hydrometer to the interested scale marks. In this paper, hydrometer calibration set up of UME is introduced and obtained measurement results from this set up are given for a sample hydrometer.
机译:液体比重计校准可以通过“ Cuckow方法”执行,其中仅使用一种液体就足以校准整个秤。用这种方法先在空气中然后在参考液体中称重比重计。参考液体的密度是众所周知的,并且可以通过使用相关公式中的称重值和液体的密度值来确定刻度刻度的相关密度值。通过将比重计浸入感兴趣的刻度标记,可以用这种方式校准比重计的整个刻度。本文介绍了UME的比重计校准设置,并从该设置中获得了样品比重计的测量结果。