Characterizing the Electron Response and Position Sensitivity for Radiation in Plastic Scintillators

机译:表征塑料闪烁体中辐射的电子响应和位置灵敏度

获取原文

摘要

To develop a tissue-equivalent dosimeter based on LET measurement, we examine the characteristics of a candidate material, plastic scintillators. The energy resolutions of plastic scintillators are examined by using the measured electron responses defined as the relative light output per unit energy of three plastic scintillators EJ-200, EJ-212, and EJ-252. The position sensitivity characteristic is examined by using a square-aligned plastic scintillator rod, and multi-segmented photomultiplier tubes are attached to both ends of the respective rod. The relation between signals from both end PMTs and the position of radiation incidence in the rod are obtained.
机译:为了开发基于LET测量的组织等效剂量计,我们检查了候选材料塑料闪烁体的特性。通过使用测量的电子响应来检查塑料闪烁体的能量分辨率,该电子响应定义为三个塑料闪烁体EJ-200,EJ-212和EJ-252的每单位能量的相对光输出。通过使用正方形对准的塑料闪烁器棒检查位置灵敏度特性,并将多段光电倍增管连接到相应棒的两端。获得了来自两个端部PMT的信号与棒中辐射入射位置之间的关系。