首页> 外文会议>Conference on radiation detectors in medicine, industry, and national security >Measurement of X-ray spectra using a recent YAP(Ce)-MPPC detector
【24h】

Measurement of X-ray spectra using a recent YAP(Ce)-MPPC detector

机译:使用最新的YAP(Ce)-MPPC检测器测量X射线光谱

获取原文
获取原文并翻译 | 示例

摘要

To measure X-ray spectra with high count rates, we developed a detector consisting of a cerium-doped yttrium aluminum perovskite [YAP(Ce)] crystal and a recent multipixel photon counter (MPPC). Scintillation photons are detected using the MPPC, and the photocurrents flowing through the MPPC are converted into voltages and amplified using a high-speed current-voltage (Ⅰ-Ⅴ) amplifier. The MPPC bias voltage was set to a value at the pre-Geiger mode to perform zero-dark counting. The event-pulse widths were approximately 200 ns, and the widths were extend to approximately 1 μs. X-ray spectra were measured using a multichannel analyzer (MCA) for pulse-height analysis. The photon energy was roughly determined by the two-point calibration using tungsten K photons and iodine K fluorescence. Using the YAP(Ce)-MPPC detector, first-generation dual-energy computed tomography was accomplished using iodine and gadolinium contrast media.
机译:为了测量高计数率的X射线光谱,我们开发了一种探测器,该探测器由掺铈的钇铝钛矿[YAP(Ce)]晶体和最近的多像素光子计数器(MPPC)组成。使用MPPC检测闪烁的光子,然后将流过MPPC的光电流转换为电压,并使用高速电流-电压(I-Ⅴ)放大器进行放大。在预盖革模式下将MPPC偏置电压设置为某个值,以执行零暗计数。事件脉冲宽度大约为200 ns,并且宽度扩展到大约1μs。使用多通道分析仪(MCA)测量X射线光谱以进行脉冲高度分析。通过使用钨K光子和碘K荧光的两点校准来大致确定光子能量。使用YAP(Ce)-MPPC检测器,使用碘和g造影剂完成了第一代双能计算机断层扫描。

著录项

  • 来源
  • 会议地点 San Diego(US)
  • 作者单位

    Department of Physics, Iwate Medical University, 2-1-1 Nishitokuta, Yahaba, Iwate 028-3694,Japan;

    Department of Physics, Iwate Medical University, 2-1-1 Nishitokuta, Yahaba, Iwate 028-3694,Japan;

    Department of Radiology, School of Medicine, Iwate Medical University, 19-1 Uchimaru,Morioka, Iwate 020-0023, Japan;

    Department of Radiology, School of Medicine, Iwate Medical University, 19-1 Uchimaru,Morioka, Iwate 020-0023, Japan;

    Central Radiation Department, Iwate Medical University Hospital, 19-1 Uchimaru, Morioka, Iwate 020-0023, Japan;

    Department of Surgery, Toho University Ohashi Medical Center, 2-17-6 Ohashi, Meguro, Tokyo 153-8515, Japan;

    Department of Surgery, Toho University Ohashi Medical Center, 2-17-6 Ohashi, Meguro, Tokyo 153-8515, Japan;

    Department of Surgery, Toho University Ohashi Medical Center, 2-17-6 Ohashi, Meguro, Tokyo 153-8515, Japan;

    Department of Surgery, Toho University Ohashi Medical Center, 2-17-6 Ohashi, Meguro, Tokyo 153-8515, Japan;

    Department of Surgery, Toho University Ohashi Medical Center, 2-17-6 Ohashi, Meguro, Tokyo 153-8515, Japan;

    Department of Surgery, Toho University Ohashi Medical Center, 2-17-6 Ohashi, Meguro, Tokyo 153-8515, Japan;

  • 会议组织
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

    YAP(Ce)-MPPC detector; high-speed Ⅰ-Ⅴ amplifier; zero dark counting; X-ray spectra; pre-Geiger mode; high count rate; dual-energy CT;

    机译:YAP(Ce)-MPPC检测器;高速Ⅰ-Ⅴ放大器;零暗计数; X射线光谱;预盖革模式高计数率;双能CT;