Quasi-monochromatic parallel radiography achieved with a plane-focus x-ray tube

机译:平面聚焦X射线管实现的准单色平行射线照相

获取原文
获取原文并翻译 | 示例

摘要

Fundamental study on quasi-monochromatic parallel radiography using a polycapillary plate and a plane-focus x-ray tube is described. The x-ray generator consists of a negative high-voltage power supply, a filament (hot cathode) power supply, and an x-ray tube. The negative high-voltage is applied to the cathode electrode, and the transmission type target (anode) is connected to the ground potential. The maximum voltage and current of the power supply were -100kV (peak value) and 3.0mA, respectively. In this experiment, the tube voltage was regulated from 20 to 25kV, and the tube current was regulated by the filament temperature and ranged from 1.0 to 3.0mA. The exposure time is controlled in order to obtain optimum film density, and the focal spot diameter was about 10mm. The polycapillary plate is J5022-21 made by Hamamatsu Photonics Inc., and the outside and effective diameters are 87 and 77mm, respectively. The thickness and the hole diameter of the polycapillary are 1.0mm and 25 μm, respectively. The x-rays from the tube are formed into parallel beam by the polycapillary, and the radiogram is taken using an industrial x-ray film of Fuji IX 100 without using a screen. In the measurement of image resolution, we employed three brass spacers of 2, 30, and 60mm in height. By the test chart, the resolution fell according to increases in the spacer height without using a polycapillary. In contrast, the resolution slightly fell with corresponding increases in the height by the polycapillary. In angiography, fine blood vessels of about 100 μm are clearly visible.
机译:描述了使用多毛细管板和平面聚焦X射线管进行准单色平行射线照相的基础研究。 X射线发生器由负高压电源,灯丝(热阴极)电源和X射线管组成。负高压施加到阴极,透射型靶(阳极)连接到地电位。电源的最大电压和电流分别为-100kV(峰值)和3.0mA。在该实验中,管电压调节在20至25kV,管电流通过灯丝温度调节,范围在1.0至3.0mA。为了获得最佳的膜密度,控制曝光时间,并且焦点直径为约10mm。多毛细管板是由Hamamatsu Photonics Inc.制造的J5022-21,其外径和有效直径分别为87mm和77mm。多毛细管的厚度和孔径分别为1.0mm和25μm。来自管的x射线被多毛细管形成平行光束,并且使用不带屏幕的Fuji IX 100工业x射线胶片来拍摄放射线照片。在测量图像分辨率时,我们使用了3个高度分别为2、30和60mm的黄铜垫片。根据测试图,不使用多毛细管,分辨率会随着间隔物高度的增加而降低。相反,分辨率随着多毛细管高度的增加而略有下降。在血管造影中,清晰可见约100μm的细血管。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号