Image guided breast biopsy using 3D ultrasound and stereotactic mammography

机译:使用3D超声和立体定位乳腺摄影术的图像引导式乳腺活检

获取原文
获取原文并翻译 | 示例

摘要

A 3D ultrasound (US)-guided biopsy system was developed to supplement stereotactic mammography (SM) with near real-time 3D and real-time 2D US imaging. We have combined features from SM and US guided biopsy, including breast stabilisation, a confined needle trajectory and dual modality imaging. We have evaluated our procedure using breast phantoms, in terms of its accuracy with US-guided biopsy. Phantoms made of animal tissue with embedded phantom 'lesions' allowed us to test the biopsy accuracy of our procedure. We have also registered the SM image space to US image space, and both spaces to the mechanical geometry of the needle trajectory. Evaluation experiments have shown that our US-guided biopsy procedure was capable of placing the needle tip with 0.85 mm accuracy at a target identified in the 3D image. We also identified that we could successfully biopsy artificial lesions that were 3.2 mm in diameter, with a 96% success rate. As an adjunct to stereotactic mammography, we propose that this system could provide more complete information for target identification and real-time monitoring of needle insertion, as well as providing a means for rapid confirmation of biopsy success with 3D ultrasound.
机译:开发了一种3D超声(US)引导的活检系统,以通过近实时3D和实时2D US成像来补充立体定向乳腺摄影(SM)。我们结合了SM和美国引导下的活检的功能,包括乳房稳定,狭窄的针头轨迹和双模式成像。我们已经评估了使用乳房幻影术进行手术的准确性,以及在美国指导下进行活检的准确性。由动物组织制成的带有幻影“病变”的幻影使我们能够测试手术活检的准确性。我们还将SM图像空间注册为US图像空间,并将两个空间注册为针迹的机械几何形状。评估实验表明,我们的US引导活检程序能够将针尖以0.85 mm的精度放置在3D图像中确定的目标上。我们还确定,我们可以成功地对直径为3.2毫米的人工病变进行活检,成功率为96%。作为立体定位乳房X线摄影术的附件,我们建议该系统可以为目标识别和针头插入的实时监控提供更完整的信息,并提供一种快速确认3D超声活检成功的方法。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号