Accurate Measurement of Monochrome Luminance Palettes for the Calibration of Medical LCD Monitors

机译:单色亮度调色板的准确测量,用于医用LCD显示器的校准

获取原文
获取原文并翻译 | 示例

摘要

Grayscale medical monitors are commonly calibrated by transforming the image display values sent to a graphic controller using a lookup table (LUT). The calibration LUT is deduced from the uncalibrated luminance response (uLR) of the display system. The uLR of liquid crystal display (LCD) systems is poorly behaved with significant discontinuities occurring in the relative luminance changes. Accurate grayscale calibration of LCD devices thus requires a measurement of the luminance for the full palette of possible output values. A method is reported to acquire the uLR of LCD displays, generate a LUT to achieve precise calibration, and assess the accuracy of the calibration results. A palette of 766 luminance values can be measured in 12 minutes. The accuracy of the method permits the evaluation of relative luminance changes, dL/L, to be made with a precision of .0002 to .0007 for luminance values between 1000 and 1 cd/m~2. For seven LCD monitors, 766 values for the uLR were measured and calibration tables deduced. The calibrated luminance response (cLR) for 256 gray values was then compared to the DICOM standard. The root mean squared error of the observed JNDs per luminance interval values ranged from .37 to .59 which is less than the AAPM recommended value of 1.0. A full calibration of this type should be done at installation. However, the stability of LCD systems suggests that periodic recalibration will not be necessary.
机译:灰度医疗监视器通常通过使用查找表(LUT)转换发送到图形控制器的图像显示值进行校准。根据显示系统的未校准亮度响应(uLR)推导校准LUT。液晶显示器(LCD)系统的uLR表现不佳,相对亮度变化中出现明显的不连续性。因此,LCD设备的精确灰度校准需要对整个可能的输出值调色板进行亮度测量。据报道,有一种方法可以获取LCD显示器的uLR,生成LUT以实现精确校准,并评估校准结果的准确性。可以在12分钟内测量766个亮度值的调色板。该方法的精度允许对亮度值在1000到1 cd / m〜2之间的相对亮度变化dL / L进行评估,精度为.0002到.0007。对于七个LCD监视器,测量了766个uLR值并推导出校准表。然后将256个灰度值的校准亮度响应(cLR)与DICOM标准进行比较。每个亮度间隔值观测到的JND的均方根误差范围为0.37至0.59,小于APM推荐值1.0。此类型的完整校准应在安装时进行。但是,LCD系统的稳定性表明,不需要定期进行重新校准。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号