首页> 外文会议>Clinical and biomedical spectroscopy >Using broadband spatially resolved NIRS to assess muscle oxygenation during altered running protocols
【24h】

Using broadband spatially resolved NIRS to assess muscle oxygenation during altered running protocols

机译:使用宽带空间分辨NIRS评估跑步方案变更后的肌肉氧合作用

获取原文
获取原文并翻译 | 示例

摘要

We used spatially resolved near-infrared spectroscopy (SRS-NIRS) to assess calf and thigh muscle oxygenation during running on a motor-driven treadmill. Two protocols were used: An incremental speed protocol (velocity = 6-12 km/h, Av = 2 km/h) was performed in 3 minute stages, while a pacing paradigm modulated step frequency alternatively (2.3 Hz [S_(Low)]; 3.3 Hz [S_(High)]) during a constant velocity for 2 minutes each. A SRS-NIRS broadband system (600 - 1000 nm) was used to measure total haemoglobin concentration and oxygen saturation (SO_2). An accelerometer was placed on the hip joints to measure limb acceleration through the experiment. The data showed that the calf (SO_2 58 to 42%) desaturated to a significantly lower level than the thigh (61 to 54%). During the pacing protocol, SO_2 was significantly different between the S_(low) vs- S_(high) trials. Additionally, physiological data as measured by spirometry were different between the S_(low) vs. S_(High) pacing trials (VO2 (2563± 586 vs. 2503 ± 605 Ml/min). Significant differences in VO2 at the same workload (speed) indicate alterations in mechanical efficiency. These data suggest that SRS broadband NIRS can be used to discern small changes in muscle oxygenation, making this device useful for metabolic exercise studies in addition to spirometry and movement monitoring by accelerometers.
机译:我们使用空间分辨近红外光谱(SRS-NIRS)来评估在电动跑步机上跑步时小腿和大腿的氧合。使用了两种协议:以增量速度协议(速度= 6-12 km / h,Av = 2 km / h)在3分钟内执行,而起搏范式交替调制步进频率(2.3 Hz [S_(Low)] ; 3.3 Hz [S_(High)])在恒定速度下持续2分钟。使用SRS-NIRS宽带系统(600-1000 nm)测量总血红蛋白浓度和氧饱和度(SO_2)。将加速度计放在髋关节上,以通过实验测量四肢的加速度。数据表明,小腿(SO_2为58至42%)的去饱和度大大低于大腿(61至54%)。在起搏方案中,S_2(低)vs-S_(高)试验之间的SO_2明显不同。此外,通过肺活量测定法测量的生理数据在S_(低)与S_(高)起搏试验之间是不同的(VO2(2563±586 vs. 2503±605 Ml / min)。在相同工作量(速度)下,VO2的显着差异这些数据表明,SRS宽带NIRS可用于识别肌肉氧合作用的细微变化,从而使该设备不仅可用于肺活量测定和加速度计的运动监测,还可用于代谢运动研究。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号