首页> 外文会议>Chemical Sensors 10 and MEMS/NEMS 10 >Preparation of Fine Implantable Needle Type Biosensors for Blood Vessel Glucose Monitoring
【24h】

Preparation of Fine Implantable Needle Type Biosensors for Blood Vessel Glucose Monitoring

机译:用于血管葡萄糖监测的精细植入式针式生物传感器的制备

获取原文
获取原文并翻译 | 示例

摘要

Fine needle type glucose sensors with variety of outer membrane were prepared and their glucose sensor properties were evaluated by in vitro measurement in phosphate buffer solution and horse serum, and in vivo measurement using the veins of rats and rabbits. Extra-outer membrane materials of polyglutamic acid (PGA) and heparin, which provide hydrophilic surface were introduced in addition to polyurethane/polydimethylsiloxane composite outer membrane. Glucose sensors without extra-outer film and with PGA extra-outer film showed glucose response on the day of implantation, while clear response was not obtained on the second day of implantation due to the adsorption of thrombus on the sensor surface. On the other hand, the sensor with heparin outer membrane provided satisfactory glucose response also on the second day.
机译:制备了具有多种外膜的细针型葡萄糖传感器,并通过在磷酸盐缓冲液和马血清中的体外测量以及使用大鼠和兔子的静脉的体内测量来评估其葡萄糖传感器的性能。除了聚氨酯/聚二甲基硅氧烷复合材料外膜外,还引入了提供亲水性表面的聚谷氨酸(PGA)和肝素的外膜材料。没有多余外膜和带有PGA外面膜的葡萄糖传感器在植入当天显示葡萄糖反应,而在植入第二天由于血栓在传感器表面的吸附而未获得清晰的反应。另一方面,带有肝素外膜的传感器在第二天也提供了令人满意的葡萄糖反应。

著录项

  • 来源
  • 会议地点 Honolulu HI(US)
  • 作者

    K. Edagawa; M. Yasuzawa;

  • 作者单位

    Department of Chemical Science and Technology, The University of Tokushima, Tokushima 770-8506, JAPAN;

    Department of Chemical Science and Technology, The University of Tokushima, Tokushima 770-8506, JAPAN;

  • 会议组织
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

  • 入库时间 2022-08-26 14:19:48

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号