首页> 外文会议>Bio-MEMS and medical microdevices >Optimization of dielectrophoretic separation and concentration of pathogens in complex biological samples
【24h】

Optimization of dielectrophoretic separation and concentration of pathogens in complex biological samples

机译:复杂生物样品中电泳分离和病原体浓度的优化

获取原文
获取原文并翻译 | 示例

摘要

Sample preparation is a key issue of modern analytical methods for in vitro diagnostics of diseases with microbiological origins: methods to separate bacteria from other elements of the complex biological samples are of great importance. In the present study, we investigated the DEP force as a way to perform such a de-complexification of the sample by extracting micro-organisms from a complex biological sample under a highly non-uniform electric field in a micro-system based on an interdigitated electrodes array. Different parameters were investigated to optimize the capture efficiency, such as the size of the gap between the electrodes and the height of the capture channel. These parameters are decisive for the distribution of the electric field inside the separation chamber. To optimize these relevant parameters, we performed numerical simulations using COMSOL Multiphysics and correlated them with experimental results. The optimization of the capture efficiency of the device has first been tested on micro-organisms solution but was also investigated on human blood samples spiked with micro-organisms, thereby mimicking real biological samples.
机译:样品制备是现代分析方法的一个关键问题,用于对具有微生物起源的疾病进行体外诊断:将细菌与复杂生物样品中其他元素分离的方法非常重要。在本研究中,我们研究了DEP力,该方法是通过在基于交叉指的微系统中的高度非均匀电场下从复杂生物样品中提取微生物来执行样品的这种复杂化操作电极阵列。研究了不同的参数以优化捕获效率,例如电极之间间隙的大小和捕获通道的高度。这些参数对于分离室内的电场分布起决定性作用。为了优化这些相关参数,我们使用COMSOL Multiphysics进行了数值模拟,并将其与实验结果相关联。该设备的捕获效率的优化首先在微生物溶液中进行了测试,但也对掺有微生物的人体血液样本进行了研究,从而模仿了真实的生物样本。

著录项

  • 来源
    《Bio-MEMS and medical microdevices》|2013年|876507.1-876507.10|共10页
  • 会议地点 Grenoble(FR)
  • 作者单位

    Department of microTechnology for Biology and Health, CEA LETI-Minatec, 17 rue des Martyrs, 38054 Grenoble Cedex 9, France;

    Department of microTechnology for Biology and Health, CEA LETI-Minatec, 17 rue des Martyrs, 38054 Grenoble Cedex 9, France;

    BioMerieux, Technology Research Department, 5 rue des Berges, 38000 Grenoble, France;

    Department of microTechnology for Biology and Health, CEA LETI-Minatec, 17 rue des Martyrs, 38054 Grenoble Cedex 9, France;

    SATIE, CNRS, ENS Cachan, 61 av. President Wilson, 94235 Cachan Cedex, France;

    SATIE, CNRS, ENS Cachan, 61 av. President Wilson, 94235 Cachan Cedex, France;

  • 会议组织
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

    sample preparation; blood; sepsis; dielectrophoresis (DEP); cell separation; lab-on-a-chip;

    机译:样品制备;血液;败血症介电电泳(DEP);细胞分离;芯片实验室;