首页> 外文会议>APEMC2010;Asia-Pacific Symposium on Electromagnetic Compatibility >Calibration measurement setup for band-selective personal exposure meters
【24h】

Calibration measurement setup for band-selective personal exposure meters

机译:波段选择个人曝光仪的校准测量设置

获取原文

摘要

In this paper a calibration measurement setup for band-selective personal exposure meters is presented. The equipment and the methods of the measurement procedure are described together with the uncertainty assessment of the measurement equipment, the calibration and the measurement procedure. The average measurement error for the reference measurement is ± 0.32 dB, while the expanded uncertainty with a confidence interval of 95% is calculated to 2.5 dB. Furthermore a straight forward calibration method is described and tested using the EME SPY 120 device from Antenessa. Up to seven different measurements must be performed for a full characterization of personal exposure meters. The presented measurement setup leads to a higher accuracy of the calibration and the achieved results will allow for an improved exposure assessment in experimental and epidemiological studies.
机译:在本文中,介绍了用于波段选择性个人曝光仪的校准测量设置。描述了测量过程的设备和方法,以及测量设备的不确定度评估,校准和测量过程。参考测量的平均测量误差为±0.32 dB,而置信区间为95%的扩展不确定度计算为2.5 dB。此外,还介绍了一种简单的校准方法,并使用了Antenessa的EME SPY 120设备进行了测试。为了充分表征个人测光表,最多必须执行七个不同的测量。提出的测量设置可导致更高的校准精度,并且所获得的结果将有助于改进实验和流行病学研究中的暴露评估。