Differential Scanning Calorimetry (DSC) measurements are routinely used to determine enthalpies of phase change, phase transition temperatures, glass transition temperatures, and heat capacities. In order to obtain data on the amount of phases during phase change, time-temperature lags, which are inherent to the measurement process, must be estimated through a computational analysis. An analytical model is proposed for the systematic error of the instrument. Numerical simulation results are compared against experimental data obtained at different heating and cooling rates.
机译:差示扫描量热法(DSC)测量通常用于确定相变,相变温度,玻璃化转变温度和热容量的焓。为了获得相变过程中的相数数据,必须通过计算分析来估算测量过程中固有的时间-温度滞后。针对仪器的系统误差,提出了一种解析模型。将数值模拟结果与在不同加热和冷却速率下获得的实验数据进行比较。