【24h】

Analysis of strain underestimation due to pulse base in a BOTDA sensor system

机译:BOTDA传感器系统中由于脉冲基引起的应变低估分析

获取原文
获取原文并翻译 | 示例

摘要

This article presents a pulse base effect on the Brillouin gain spectrum that can be obtained by BOTDA, which uses a pump-probe system. For theoretical analysis of the pulse base effect, we propose a theoretical model in which the interaction between the pulse base and probe light is taken into consideration. Strain underestimation caused by the pulse base is demonstrated by experiments and theoretical simulations.
机译:本文介绍了对布里渊增益频谱的脉冲基效应,该效应可以通过使用泵浦探针系统的BOTDA获得。为了对脉冲基效应进行理论分析,我们提出了一个理论模型,其中考虑了脉冲基与探测光之间的相互作用。实验和理论模拟证明了脉冲基引起的应变低估。