首页> 外文会议>Annual Conference of Japanese Society for Medical and Biological Engineering;Annual International Conference of the IEEE Engineering in Medicine and Biology Society >Temperature distributions measurement of high intensity focused ultrasound using a thin-film thermocouple array and estimation of thermal error caused by viscous heating
【24h】

Temperature distributions measurement of high intensity focused ultrasound using a thin-film thermocouple array and estimation of thermal error caused by viscous heating

机译:使用薄膜热电偶阵列测量高强度聚焦超声的温度分布并估算由粘性加热引起的热误差

获取原文
获取原文并翻译 | 示例

摘要

To improve the throughput of high intensity focused ultrasound (HIFU) treatment, we have considered a focus switching method at two points. For this method, it is necessary to evaluate the thermal distribution under exposure to ultrasound. The thermal distribution was measured using a prototype thin-film thermocouple array, which has the advantage of minimizing the influence of the thermocouple on the acoustic and temperature fields. Focus switching was employed to enlarge the area of temperature increase and evaluate the proposed evaluation parameters with respect to safety and uniformity. The results indicate that focus switching can effectively expand the thermal lesion while maintaining a steep thermal boundary. In addition, the influence caused by the thin-film thermocouple array was estimated experimentally. This thermocouple was demonstrated to be an effective tool for the measurement of temperature distributions induced by HIFU.
机译:为了提高高强度聚焦超声(HIFU)治疗的通量,我们考虑了两种方法的焦点切换方法。对于此方法,有必要评估暴露于超声下的热分布。使用原型薄膜热电偶阵列测量热分布,其优点是最大程度地减少了热电偶对声场和温度场的影响。使用焦点切换来扩大温度升高的区域,并就安全性和均匀性评估建议的评估参数。结果表明,焦点切换可以在保持陡峭的热边界的同时有效扩展热损伤。另外,还通过实验估算了薄膜热电偶阵列引起的影响。该热电偶被证明是测量HIFU引起的温度分布的有效工具。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号