首页> 外文会议>IEEE Micro Electro Mechanical Systems Conference >A calorimetric flow sensor for ultra-low flow applications using electrochemical impedance
【24h】

A calorimetric flow sensor for ultra-low flow applications using electrochemical impedance

机译:使用电化学阻抗的超低流量应用量热流量传感器

获取原文
获取原文并翻译 | 示例

摘要

We present a calorimetric flow sensor which utilizes electrochemical impedance (EI) measurement to sense flow-mediated heat transfer. Sensors were fabricated from thin-film platinum and Parylene C, rendering them biocompatible, flexible, and corrosion-resistant. A novel sensing method based on asymmetric impedance dips during heater pulsing allowed flow measurement at ultra-low flow velocities (less than 200 μm/s), and simultaneous EI measurement from multiple electrode pairs using a custom multiplexer setup enabled highly sensitive, bidirectional flow transduction. Sensors tested in phosphate-buffered saline achieved a 2σ resolution of 19.1 μm/s, which represents a greater than 4x improvement over previously reported EI-based sensors. The sensor's biocompatible and flexible construction is ideal for microfluidics, lab-on-chip, and in vivo applications.
机译:我们提出了一种利用电化学阻抗(EI)测量以感测流介导的热传递的量热流量传感器。传感器由薄膜铂和聚对二甲苯C制成,使其具有生物相容性,柔性和耐腐蚀性能。一种基于加热器脉冲过程中不对称阻抗骤降的新颖传感方法,可在超低流速(小于200μm/ s)下进行流量测量,并使用定制的多路复用器设置同时从多个电极对进行EI测量,从而实现了高度灵敏的双向流量转换。在磷酸盐缓冲液中测试过的传感器的2σ分辨率为19.1μm/ s,这比以前报道的基于EI的传感器提高了4倍以上。该传感器具有生物相容性和柔性结构,非常适合微流控,芯片实验室和体内应用。