首页> 外文会议>2015 International Conference on Industrial Instrumentation and Control >Evanescent absorption based refractive index measurement at 1.3 µm
【24h】

Evanescent absorption based refractive index measurement at 1.3 µm

机译:基于逝吸收的折射率测量,为1.3 µm

获取原文
获取原文并翻译 | 示例

摘要

In this article the different works of optical fiber based refractive index measurements is reviewed. An experimental study and analysis of fiber optic cable based refractive index sensor has been carried out to measure the change in signal for different solutions of known refractive indices. The model is based on evanescent field absorption in medium of different refractive indices. A multi-mode step index fiber of diameter 62.5/125±5 μm is used to operate at near infrared (@1310 nm) laser source. The cladding of the fiber was removed to a maximum of 40% of its thickness and the fiber was dipped in water, ethanol and chloroform. The output was measured using a digital power meter. Maximum deviation of 0.003 is obtained in the linear fit of the normalized signal with known refractive index values.
机译:本文回顾了基于光纤的折射率测量的不同工作。已经进行了基于光纤电缆的折射率传感器的实验研究和分析,以测量已知折射率的不同解决方案的信号变化。该模型基于在不同折射率的介质中的van逝场吸收。直径为62.5 / 125±5μm的多模阶跃折射率光纤用于在近红外(@ 1310 nm)激光源下工作。除去纤维的包层至最大厚度的40%,并将纤维浸入水,乙醇和氯仿中。使用数字功率计测量输出。在具有已知折射率值的归一化信号的线性拟合中获得0.003的最大偏差。