首页> 外文会议>2015 41st Annual Northeast Biomedical Engineering Conference >Simulation study for the design of an EIT system for cardiac output monitoring
【24h】

Simulation study for the design of an EIT system for cardiac output monitoring

机译:用于心输出量监测的EIT系统设计的仿真研究

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

This simulation study explores system design requirements needed for an EIT system that will be capable of accurately monitoring cadiac output. The 4D XCAT model is used to construct a realastic time-depedent conductivity profile of the chest. Forward and inverse simulations are calculated using Dartmouth's NDRM EIT software package. Error analysis revealed that an SNR of 85-91 dB is preferred to confidently measure cardiac output.