Remote patient monitoring (RPM) provides an efficient method to improve patient care in hospitals. Zigbee network technology with low-power consuming sensors has the capability to provide enhanced mobility and comfort to patients. A single Zigbee based sensor network is unable supporting all the patients on a general ward. The allocation of a ZigBee sensor network to each patient can address this limitation. However a large number of ZigBee networks can generate channel overlap problems due to the limited number of channels used by ZigBee. In the presence of channel overlap, multiple Zigbee networks may experience problems when handling continuous transmission of vital signs in RPM. However the adoption of appropriate time intervals for transmission of vital signs from sensors can rectify this problem thereby providing reliable transmission of vital signs.
机译:远程病人监护(RPM)提供了一种有效的方法来改善医院的病人护理。具有低功耗传感器的Zigbee网络技术能够为患者提供增强的移动性和舒适度。一个基于Zigbee的传感器网络无法支持普通病房中的所有患者。将ZigBee传感器网络分配给每个患者可以解决此限制。但是,由于ZigBee使用的信道数量有限,大量ZigBee网络可能会产生信道重叠问题。在通道重叠的情况下,当处理RPM中生命体征的连续传输时,多个Zigbee网络可能会遇到问题。但是,采用适当的时间间隔从传感器传输生命体征可以纠正此问题,从而提供生命体征的可靠传输。