首页> 外文会议>2010 IEEE International Workshop on Medical Measurements and Applications Proceedings >The effect of blood pressure variability on the estimation of the systolic and diastolic pressures
【24h】

The effect of blood pressure variability on the estimation of the systolic and diastolic pressures

机译:血压变异性对收缩压和舒张压估计的影响

获取原文
获取原文并翻译 | 示例

摘要

Systolic (SBP) and diastolic (DBP) blood pressure are two parameters that are widely used in clinical and personal health monitoring and that are subject to significant physiological variability. The estimation of SBP and DBP is therefore subject to two sources of uncertainty: 1) measurement error, and 2) physiological variability. According to the ANSI/AAMI SP10 standard, the measurement error of an automated sphygmomanometer is limited to ±5 mmHg relative to simultaneous reference readings by at least two trained observers employing a calibrated manometer. In order to directly compare the contributions of physiological variability and measurement error to SBP and DBP estimation uncertainty, we analyzed continuous short recordings of arterial pulses (≪ 5 min) and calculated the percentage of beats in which the SBP and DBP exceeds ±5 mmHg relative to the mean value in the analyzed interval. In a group of ICU patients, this percentage was 33.0% for SBP and 14.9% for DBP on average, while in a group of healthy individuals this percentage was 27.5% for SBP and 17.8% for DBP on average. True outliers occurred in 5.17% of the SBP and 12.2% of the DBP values in ICU patients, and 3.27% of the SBP and 3.63% of the DBP values in healthy individuals. These results indicate that blood pressure variability is an important contributor to SBP and DBP estimation uncertainty, and devices that intelligently determine the best time to initiate a recording and the frequency of recordings may alleviate this uncertainty.
机译:收缩压(SBP)和舒张压(DBP)是在临床和个人健康监测中广泛使用的两个参数,它们具有明显的生理变异性。因此,对SBP和DBP的估计有两个不确定性来源:1)测量误差,以及2)生理变异性。根据ANSI / AAMI SP10标准,相对于至少两个训练有素的压力计的受过训练的观察者的同时参考读数,自动血压计的测量误差限制为±5 mmHg。为了直接比较生理变异性和测量误差对SBP和DBP估计不确定性的贡献,我们分析了连续短时记录的动脉搏动(≪ 5分钟),并计算了相对压力(SBP)和DBP超过±5 mmHg的搏动百分比到分析间隔中的平均值。在一组ICU患者中,SBP的平均百分比为33.0%,DBP的平均百分比为14.9%,而健康个体组的SBP的平均百分比为27.5%,DBP的平均百分比为17.8%。在ICU患者中,真实异常值发生在SBP的5.17%和DBP值的12.2%,在健康个体中发生在SBP的3.27%和DBP值的3.63%。这些结果表明,血压变异性是SBP和DBP估计不确定性的重要因素,智能地确定最佳启动记录时间和记录频率的设备可以减轻这种不确定性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号