首页> 外国专利> Measurement of optical fields, e.g. above and below open waters, using spectral radiometer, is accompanied by correction of thermally-generated signal components

Measurement of optical fields, e.g. above and below open waters, using spectral radiometer, is accompanied by correction of thermally-generated signal components

机译:光场的测量,例如使用光谱辐射计,在开阔水域的上方和下方,对热产生的信号分量进行校正

摘要

To correct thermally-generated signal components of a photosensor in the spectral radiometer before and after measurement of light (or darkness) intensity, signals are received over equal time intervals. The mean value of the readings is subtracted from the brightness or darkness measurement. Preferred Features: Measurements are transmitted from the spectrometer slit aperture by spacer-separated optical conductors. Spectral strips are produced in the focal plane, separated by dark gaps caused by the spacers. A CCD detector is employed. To correct for optical cross-interference and scattered light, the signal profile in dark regions adjacent to spectral strips (measurement channel) is measured and used to determine a spectral profile of signal patterns to be subtracted. Amplitude of the signal pattern is defined in spectral strip wavelength intervals, in which cross-interference and scattered signals are detected, but not external light. With similar signal profiles light signals received by various light measurement units are used for correction of the cross-interference in dark regions at upper edges, lower edges and the central region of the CCD sensor. For radiometric calibration of the multi-beam spectrometer using a number of light measurement sensors, a reference standard is used. A calibration correction is carried out from actual measurements and post-calibration measurements, producing a ratio Rk/RN. The correction strongly smoothes out additional noise. Variable optical attenuators are employed to standardize signal strengths detected by differing detectors, and are used in conjunction with a reference standard to match the signals. Optical fields may be measured in diverse open waters, above and below the surface. Signals from the upper (in air) sensors are matched by variable attenuators in the optical path, with the signals from underwater sensors. Calibration of attenuated spectra are made from transmission characteristics of the attenuation sections. For standardization of attenuated spectra an additional light measurement unit with its own detection unit is employed. This receives an incident light intensity in a narrow wavelength interval. Its signal is used for automatic control of the attenuators. For automatic adjustment of an illumination interval at given measurement depth during reception of a time-integrated signal, an additional, flashing measurement unit above to the water surface, with its own detector unit, is employed. To determine absorption coefficients of an optical field with the help of spectral radiometers, a net radiation/beam intensity (E(zo), E(zu)) and a scalar radiation intensity (E(z0), E(zu)) are determined simultaneously in two planes, an upper plane zo and a lower plane zu. From the values E(zo), E(zu) the attenuation coefficient K of the net radiation intensity is calculated as K(zm) = DELTA z-1 . ln(E(zo)/E(zu)). zm is the center of the zo and zu planes and DELTA z is their separation. An additional expression in the form of the Gershun equation is provided, for solution. Net radiation intensities and scalar radiation intensities are determined successively with two light measurement units having a specially-designed angular sensitivity profile, exhibiting a blend of cosine- and 2 pi -reception characteristics in two planes (zo, zu).
机译:为了在测量光(或暗)强度之前和之后,在光谱辐射计中校正光传感器中由热产生的信号分量,在相等的时间间隔内接收信号。从亮度或暗度测量值中减去读数的平均值。首选功能:测量是通过分隔器分隔的光导体从光谱仪的狭缝孔传输的。光谱带在焦平面上产生,由间隔物引起的暗间隙隔开。使用CCD检测器。为了校正光学交叉干扰和散射光,测量与光谱带(测量通道)相邻的暗区中的信号轮廓,并将其用于确定要减去的信号图案的光谱轮廓。信号模式的幅度以光谱带波长间隔定义,在该间隔中检测到交叉干扰和散射信号,但未检测到外部光。在具有相似的信号轮廓的情况下,由各种测光单元接收到的光信号用于校正CCD传感器上边缘,下边缘和中央区域的暗区中的交叉干扰。为了使用多个光测量传感器对多光束光谱仪进行辐射校准,使用了参考标准。根据实际测量值和校准后测量值进行校准校正,得出比率Rk / RN。该校正可以大大消除额外的噪声。可变光衰减器用于标准化由不同检测器检测到的信号强度,并与参考标准结合使用以匹配信号。可以在表面上方和下方的各种开放水域中测量光场。来自上部(空中)传感器的信号与光路中的可变衰减器相匹配,与来自水下传感器的信号相匹配。衰减光谱的校准是根据衰减部分的透射特性进行的。为了标准化衰减光谱,使用了具有其自己的检测单元的附加光测量单元。这接收到窄波长间隔内的入射光强度。其信号用于衰减器的自动控制。为了在接收时间积分信号期间在给定的测量深度自动调整照明间隔,使用了一个附加的,闪烁的测量单元,该测量单元位于水面上方,并带有自己的检测器单元。为了借助光谱辐射计确定光场的吸收系数,确定净辐射/束强度(E(zo),E(zu))和标量辐射强度(E(z0),E(zu))同时在两个平面中,上平面zo和下平面zu。根据值E(zo),E(zu),将净辐射强度的衰减系数K计算为K(zm)= DELTA z -1。 ln(E(zo)/ E(zu))。 zm是zo和zu平面的中心,而DELTA z是它们的间隔。提供了Gershun方程形式的附加表达式用于求解。净辐射强度和标量辐射强度是通过两个具有特殊设计的角灵敏度曲线的光测量单元相继确定的,它们在两个平面(zo,zu)中均表现出余弦和2 pi接收特性的混合。

著录项

  • 公开/公告号DE10040889A1

    专利类型

  • 公开/公告日2002-03-28

    原文格式PDF

  • 申请/专利权人 DEUTSCHES ZENTRUM FUER LUFT-UND RAUMFAHRT E.V.;

    申请/专利号DE2000140889

  • 发明设计人 BOCHTER KURT;

    申请日2000-08-18

  • 分类号G01J3/28;

  • 国家 DE

  • 入库时间 2022-08-22 00:27:31

相似文献

  • 专利
  • 外文文献
  • 中文文献
获取专利

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号