New estimators and guidelines for better use of fetal heart rate estimators with Doppler ultrasound devices

机译:新的估算器和指南,以更好地结合多普勒超声设备使用胎儿心率估算器

摘要

Characterizing fetal wellbeing with a Doppler ultrasound device requires computation of a score based on fetal parameters. In order to analyze the parameters derived from the fetal heart rate correctly, an accuracy of 0.25 beats per minute is needed. Simultaneously with the lowest false negative rate and the highest sensitivity, we investigated whether various Doppler techniques ensure this accuracy. We found that the accuracy was ensured if directional Doppler signals and autocorrelation estimation were used. Our best estimator provided sensitivity of 95.5%, corresponding to an improvement of 14% compared to the standard estimator.
机译:用多普勒超声设备表征胎儿健康状况需要基于胎儿参数计算分数。为了正确分析从胎儿心率得出的参数,需要每分钟0.25次的准确性。同时以最低的假阴性率和最高的灵敏度,我们研究了各种多普勒技术是否能确保这种准确性。我们发现,如果使用定向多普勒信号和自相关估计,则可以确保准确性。我们最好的估算器提供了95.5%的灵敏度,与标准估算器相比,提高了14%。