首页> 外文期刊>Computational and mathematical methods in medicine >New Estimators and Guidelines for Better Use of Fetal Heart Rate Estimators with Doppler Ultrasound Devices
【24h】

New Estimators and Guidelines for Better Use of Fetal Heart Rate Estimators with Doppler Ultrasound Devices

机译:具有更好地利用多普勒超声装置的胎儿心率估算器的新估算器和指导方针

获取原文
           

摘要

Characterizing fetal wellbeing with a Doppler ultrasound device requires computation of a score based on fetal parameters. In order to analyze the parameters derived from the fetal heart rate correctly, an accuracy of 0.25 beats per minute is needed. Simultaneously with the lowest false negative rate and the highest sensitivity, we investigated whether various Doppler techniques ensure this accuracy. We found that the accuracy was ensured if directional Doppler signals and autocorrelation estimation were used. Our best estimator provided sensitivity of 95.5%, corresponding to an improvement of 14% compared to the standard estimator.
机译:用多普勒超声装置表征胎儿阱,需要基于胎儿参数计算得分。为了正确分析从胎儿心率正确地衍生的参数,需要每分钟0.25节拍的精度。同时具有最低的假负率和最高灵敏度,我们调查了各种多普勒技术是否确保了这种准确性。我们发现如果使用定向多普勒信号和自相关估计,则确保了精度。我们最优质的估算器提供了95.5%的敏感性,与标准估算器相比,对应于14%的提高。