机译:伽马相机性能参数的测量:第二部分。

获取原文

摘要

Methods are described for field measurements of the performance parameters of gamma cameras,also called Anger or scintillation cameras. Their system spatial resolution was measured with arrays of 57Co and 195Au line sources emitting gamma rays in the range of 70to 140KeV,and is of special interest to clinical nuclear medicine. The system sensitivity is determined with calibrated flood sources of the same two long-lived radionuclides. The temporal resolution is derived from counting-rate measurements with a two-source scatter phantom with 99mTc sources in test tubes. In clinical procedures the observed counting rates should be limited to incur counting losses of less than 25percent. Data for each of the three parameters are presented for more than 50gamma-camera models from eight manufacturers,employing their low-energy high-resolution,general purpose and high-sensitivity parallel-hole collimators.