Analysis of pre-flight modulator voltage calibration data for the Voyager plasma science experiment

机译:分析航海家等离子体科学实验的飞行前调制器电压校准数据

获取原文

摘要

The Voyager Plasma Science (PLS) modulator calibration (MVM) data analysis was undertaken in order to check the correctness of the fast A/D converter formulas that connect low voltage monitor signals (MV) with digital outputs (DN), to determine the proportionality constants between the actual modulator grid potential (V) and the monitor voltage (MV), and to establish an algorithm to link the digitized readouts (DN) with the actual grid potential (V). The analysis results are surprising in that the derived conversion constants deviate by fairly significant amounts from their nominal values. However, it must be kept in mind that the test results which were used for analysis may be very imprecise. Even if it is assumed that the test result errors are very large, they do no appear to be capable to account for all discrepancies between the theoretical expectations and the results of the analysis. Measurements with the flight spare instrument appear to be the only means of investigating these effects further.