首页> 美国政府科技报告 >Superconducting Detector Endstation for High-Resolution Energy-Dispersive SR-XRf
【24h】

Superconducting Detector Endstation for High-Resolution Energy-Dispersive SR-XRf

机译:用于高分辨率能量分散sR-XRf的超导探测器端站

获取原文

摘要

We have built a two-stage adiabatic demagnetization refrigerator (ADR) to operate cryogenic high-resolution x-ray detectors in synchrotron-based fluorescence applications. The detector is held at the end of a 40 cm cold finger that extends into a UHV sample chamber. The ADR attains a base temperature below 100 mK with about 24 hours hold time below 400 mK, and does not require pumping on the liquid He bath. We will discuss cryostat design and performance.