首页> 美国政府科技报告 >Statistical Review of Cesium Carbonate Fusion Measurements of Archived Glass Samples from the DWPF Process
【24h】

Statistical Review of Cesium Carbonate Fusion Measurements of Archived Glass Samples from the DWPF Process

机译:DWpF工艺中存档玻璃样品碳酸铯熔融测量的统计综述

获取原文

摘要

This technical report provides a statistical review of measurements that the Defense Waste Processing Facility (DWPF) Laboratory (Lab) generated by Inductively-Coupled Plasma - Atomic Emission Spectroscopy (ICP-AES) for samples of archived Slurry Mix Evaporator (SME) glass that were prepared using a cesium carbonate fusion dissolution method. Measurements were generated by both of the Lab's ICPs, which are designated as M-13 and M-14. Comparisons between the two ICPs suggest that for the ARG-1 measurements the M-14 provided lower (on average) concentration measurements for about 13 of the 16 elements reported by DWPF with Cr, Mn, Ni, and Zr being statistically lower (at a 5% significance level) for the ARG-1 samples. One of the three exceptions is Li, for which the M-14 yielded (on average) higher concentration measurements than the M-13. For the SME samples, the M-14 and M-13 measurements had statistically different (at the 5% significance level) means for B, Ca, Cr, Cu, Fe, Li, Mg, Mn, Na, Ni, Si, Ti, U, and Zr with the M- 14 being lower than the M-13 for all of these elements except for Li.

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信