Resolution Limitations and Optimization of the ITT F4157 Streak Tube Focus for 10-Ps Operation

机译:用于10-p操作的ITT F4157条纹管聚焦的分辨率限制和优化

获取原文

摘要

We studied the focusing properties of an ITT F4157 image tube that was biased at voltages far from the original design and is used for operation in an ultrafast (10ps) streak camera. Its output resolution at streak camera operating potentials was measured as a function of input slit width, incident-light wavelength, and focus-grid voltage, with results similar to those reported for the RCA C73435 streak tube. The temporal resolution is insensitive to focus-grid voltage for a narrow (50 mu m) input slit, but is very sensitive to focus-grid voltage for a wide (500 mu m) input slit. Spatial resolution is nearly independent of focus-grid voltage for values that give good temporal resolution. Both temporal and spatial resolution depend on the incident-light wavelength. Streak camera operation is simulated with a computer program that calculates photoelectron trajectories. Electron ray tracing describes the observed effects of slit width, incident-light wavelength, and focus-grid voltage on the output resolution. 5 ref., 10 figs. (ERA citation 13:049459)