In this study, scattered x-ray distributions were produced by irradiating a tissue equivalent phantom under clinical mammographic conditions by using Mo/Mo, Mo/Rh and W/Rh anode/filter combinations, for 25 and 30 kV tube voltages. Energy spectra of the scattered x-rays have been measured with a Cd0.9Zn0.1Te (CZT) detector for scattering angles between 30 degrees and 165 degrees. Measurement and correction processes have been evaluated through the comparison between the values of the half-value layer (HVL) and air kerma calculated from the corrected spectra and measured with an ionization chamber in a nonclinical x-ray system with a W/Mo anode/filter combination. The shape of the corrected x-ray spectra measured in the nonclinical system was also compared with those calculated using semi-empirical models published in the literature. Scattered x-ray spectra measured in the clinical x-ray system have been characterized through the calculation of HVL and mean photon energy. Values of the air kerma, ambient dose equivalent and effective dose have been evaluated through the corrected x-ray spectra. Mean conversion coefficients relating the air kerma to the ambient dose equivalent and to the effective dose from the scattered beams for Mo/Mo, Mo/Rh and W/Rh anode/filter combinations were also evaluated. Results show that for the scattered radiation beams the ambient dose equivalent provides an overestimate of the effective dose by a factor of about 5 in the mammography energy range. These results can be used in the control of the dose limits around a clinical unit and in the calculation of more realistic protective shielding barriers in mammography.
机译:在这项研究中,通过在25 kV和30 kV管电压下使用Mo / Mo,Mo / Rh和W / Rh阳极/滤光片组合在临床乳腺摄影条件下照射组织等效体模,产生了分散的X射线分布。已经使用Cd0.9Zn0.1Te(CZT)检测器测量了散射X射线的能谱,其散射角在30度到165度之间。通过对半值层(HVL)值和空气比释动能值进行比较,对测量和校正过程进行了评估,这些值由校正后的光谱计算出,并在带有W / Mo阳极/的非临床X射线系统中使用电离室进行测量过滤器组合。还比较了在非临床系统中测量的校正X射线光谱的形状与使用文献中发表的半经验模型计算出的形状。通过计算HVL和平均光子能量,可以表征在临床X射线系统中测量的散射X射线光谱。空气比释动能值,等效环境剂量和有效剂量已通过校正后的X射线光谱进行了评估。还评估了将空气比释动能与环境剂量当量以及来自Mo / Mo,Mo / Rh和W / Rh阳极/过滤器组合的散射束的有效剂量相关的平均转换系数。结果表明,对于散射的辐射束,等效的环境剂量在乳房X线照相术能量范围内提供了有效剂量的大约5倍的高估。这些结果可用于控制临床单位周围的剂量限值,并用于乳腺X线照相术中更现实的保护性屏蔽屏障的计算。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号